PRS-801
Resistance System Set Specifications
Unique wide range ohmmeter with data logging, calculating and
computer communication capabilities
Overview |
Specifications SPECIFICATIONS: Range:
Resistance from <0.1 (1.0E-1) Ohms to 200 Tera ohms (2.0E+14
ohms). Maximum Resistivity with S11.11 concentric ring 2.0E+15
ohms/square Test Voltages:
Automatic mode Default: <0.01 to 10 volts Variable 1.0E-1
to 1.0E+4 Ohms
Constant Voltage 10 volts ± <0.2 volts 1.0E+4 to 1.0E+6
Ohms
At 10 & 100 Volts 100 volts ± <2.0 volts 1.0E+6
to 2.0E+14 Ohms
Manual Mode: <0.01 to 10 volts Variable 1.0E-1
to 1.0E+5 Ohms
10 volts ± <0.2 volts 1.0E+3 to 1.0E+9 Ohms
100 volts ± <2.0 volts 2.0E+5 to 2.0E+14 Ohms
Accuracy:
Overall: ± <5% at ambient conditions (at 23oF and 30%
Rh).
Specific Range Tolerances:
1.0E-1 to 1.0E+1 ohms: ±5% Corrected for Test Lead Resistance
1.0E+1 to 1.0E+12 ohms: ±2.0% with 10-foot Test Leads
1.0E+12 to 2.0E+14 ohms: ±40% or ±0.25 Decade with
Grounded, Shielded Leads
Display:
Multi-function 2-5/8” x 1-5/8” Liquid Crystal Display
with 0.5" digit height. Displays 3-1/2 digits in Ohms, or
1.0EXX in exponential format. Ohms Display indicators: W, KW,
MW, GW and TW. Includes 19-segment analog scale (1-10 with 0.5
indication) with x1, x10, & x100 multipliers. Number of Data
Points in Memory (0 – 80). Automatic Electrification Time
(seconds), or Time required to Manually obtain steady state measurement.
Displays data HOLD, BATTERY status, MIN, MAX, AVG, REC and Test
Voltage (<10, 10, or 100 V)
LED Indicators:
14 Color Programmable LED’s from <103 to >1014 ohms.
Colors (RED, GREEN, YELLOW, or Blank/OFF) may be operator programmed
in SETUP mode.
Timer:
Times measurements In Seconds up to 99 seconds (Displayed on
LCD)
Memory:
Register stores up to 80 data points (MEM # Displayed after RESET)
RS-232 Output:
Digital format: Exponential Power followed by Integer.
Response & Electrification:
Response from >0.1 to <1.0E06 Ohms: <2.0 seconds
Average Measurement Period from 0.1 ohms to 1.0E12 Ohms 2.5 Seconds.
Calculated Electrification Period per ESD S11.11, 7.5 seconds
0.1 ohms to 10E+12 Ohms. Programmed Electrification 1.0E+12 to >2.0E14
Ohms: 15.0+ seconds
Power:
Two 9-VDC alkaline batteries. Nominal battery life 25 hours.
Dimensions:
4.0" wide x 6.0" long x 2.0" deep.
Weight:
22 ounces, with batteries
Open Circuit Current (I):
< 4 ma @ 100 Volts
Resistance Range Select:
2 Triangular button Arrows: UP and DOWN ¯. Select
Resistance
Range in single decades in Manual and Automatic/Manual modes.
Test Volts:
Manual Selection of <10, 10 or 100 volts in Manual Mode
Record/Recall:
Turns Memory Register ON if OFF. Provides access to all data
in Memory Register. Calculates and Displays Minimum, Maximum
and Average of data stored in Memory Register.
Send:
Transmits data in Memory Register to RS-232 Output Port
Clear:
Erases data in Memory Register; if in HOLD mode, discards most
recent Held Value
ON/OFF:
Power-up, perform functional & Battery tests. Power down
if ON.
Batt. Test:
Displays GOOD on LCD if acceptable voltage or Lo if unacceptable
Reset:
Enters (saves) data into Memory Register, Clears HOLD and Display.
Test:
Begins measurement sequence.
Battery Buss Cut OFF:
ON/OFF Switch isolates batteries from instrument circuits for
storage & transport
Automatic Function and Unparalleled Accuracy:
The displayed measurement is the average of eight consecutive
measurements, all within ±5% of each other. The measurement
is “saved” in the Memory Register by pressing RESET.
Data Log, Calculation & Computer
Communication:
In RECORD mode, the PRS-801 stores up to 80 measurements,
and on demand will calculate and display the Minimum, Maximum
and
Average measurements stored in the register. Using its RS-232
output and cable accessory, the PRS-801 will communicate with
computer serial ports. The instrument is supplied with the
new PROSTAT® Audit Program on a 3-1/2 inch disc, which
contains spread sheet templates for recording ESD Audit Program
Data in Excel® software. The PRS-801 computer output and
Program disk are compatible with the Semtronics SCCN® ground
monitoring system.
Technical Documents:
» PRS-801
Data Sheet |